Monochromator technology … no filters needed
The monochromator-based optical system allows for precise
selection of wavelengths from 200nm to 1000nm, with no filters required. This
range makes the SmartReader UV-Vis™ useful not only for typical ELISA ,
Bradford Lowry, and turbidity assays but also for more advanced applications
such as quantitation of nucleic acids and proteins. The SmartReader™ UV-Vis can
be set for Endpoint, Kinetic and Spectral Scanning assays.
High Throughput Capability
The MR9610 is designed for efficient processing of multiple
samples and
accepts standard 96 well plates as well as 384 well plates for high density
screening. The plate layout interface allows for a quick and intuitive setup of
a plate well’s contents. Users can highlight and label multiple wells to save
time. The display shows a graphical representation of the plate. Users can
label each well as required for: “known standard”, “unknown sample”, “blank
well” , “QC Control well”, “positive control”, “negative control”, and “empty”.
Optimized results with plate shaking and incubation
Controllable plate shaking and incubation allow for
optimization of assay results. The shaking speed can be controlled at different
intensities with options for timed and pulsed shaking before a reading, or in
between readings when performing a kinetic assay. The linear motion efficiently
and uniformly mixes all samples in both 96 and 384 well plates. Temperature can
be controlled from ambient +5C up to 45C, allowing for the processing of
enzyme kinetics, and bacterial growth assays.”
SmartDrop™ Accessory Plate
Absorbance measurements of exceptionally small volumes
(<0.5ul) can be achieved using the SmartDrop™ Accessory Plate. The
SmartDrop™ plate accommodates up to 16 samples and is ideal for checking
nucleic acid concentrations and purity for Next Gen Sequencing, or protein
concentrations.
SmartReader™ UV-Vis Microplate Reader Specifications
Model
|
Accuris
SmartReader™ UV-Vis
|
Plates Accepted
|
96-well and
384-well flat bottom
|
Display
|
10-inch color
touch screen
|
Controls
|
Touch screen
and mouse
|
Light Source
|
Xenon Flash
Lamp
|
Wavelength range
|
200-1000nm
|
Wavelength Precision
|
± 2 nm
|
Wavelength Accuracy
|
± 2 nm
|
Wavelength selection
|
Monochromator
|
Read-out range
|
0 – 4.0 OD
|
Detector
|
Dual silicon photoelectric
detectors for measurement and reference
|
Linearity @ 450nm
|
R2 ≥
0.999, [0.0 – 3.0]
|
Accuracy @ 450nm
|
1.0% + 0.003
Abs (0 – 2.0] 2.0% (2.0 – 2.5 Abs]
|
Precision@450nm
|
CV < 0.5% or
SD<0.003 (Precision mode); CV < 1.0% (Fast mode)
|
Stability@450nm
|
< 0.005A ,
(0.0 – 2.0] < 2% (2.0 – 2.5 Abs)
|
Measurement speed
|
96 well
plate: <6 seconds (fast mode); <25 seconds (precision mode) 384 well plate: <74 seconds (only precision mode available for 384)
|
Plate shaking
|
Linear; three
levels of shaking speed
|
Incubator
|
From ambient +
5°C to 45°C
|
Temperature Uniformity
|
± 0.5°C @ 37°C
|
Storage
|
16G memory,
more than 10,000 test records can be stored
|
Connection Ports
|
3 USB ports,
for PC, printer and USB-flash drive
|
Power supply
|
DC24V
6.5A
|
Dimensions (WxDxH)
|
300 x 500 x 260mm |
11.8 x 19.7 x 10.25 inches
|
Weight
|
15.5kg | 34 Lbs
|