Solid Door Built-In Vaccine Refrigerator


These premier built-in undercounter refrigerators are designed in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. Refrigerators protect pharmaceuticals at optimal temperatures, preventing waste and allowing for peak delivery.
The ADA compliant, solid door refrigerators utilize microprocessor controllers and feature temperature alarms, remote alarm contacts, and probe access ports with included probes. American Biotech Supply Vaccine Storage Refrigerators utilize HFC-free refrigerant for environmental health and energy efficiency.
 

REFRIGERATION SYSTEM


 COMPRESSOR HERMETIC, HIGH PERFORMANCE
 REFRIGERANT EPA SNAP COMPLIANT, R600A,   ISOBUTANE
 CONDENSER HYBRID FIN AND TUBE WITH LOW NOISE   FAN
 EVAPORATOR PLATE WALL
 DEFROST CYCLE OPTIMIZED, ZERO ENERGY
 
PERFORMANCE
 UNIFORMITY¹ (CABINET AIR) +/- 0.8°C
 STABILITY² (CABINET AIR) +/- 1.2°C
 MAXIMUM TEMPERATURE   VARIATION (CABINET AIR) +/- 1.4°C
 TEMPERATURE RISE AFTER AN   AFTER 8 SEC DOOR OPENINGS TEMPERATURE DID NOT EXCEED 6.4°C   AT ANY PROBE FOR ALL REQUIRED   NSF/ANSI 456 TESTING PROTOCOLS
 RECOVERY AFTER 3 MIN DOOR   OPENING ALL PROBES RECOVER TO UNDER 8°C   WITHIN 4.8 MIN.
 ENERGY CONSUMPTION 1.15 KWH/DAY
 AVERAGE HEAT REJECTION 1.57 KWH/DAY (224 BTU/H)
 NOISE PRESSURE LEVEL (DBA) 43 OR LESS INSTALLED
 PULL DOWN TIME TO NOMINAL   OPERATING TEMP 35 MIN



PERFORMANCE DATA ACQUIRED AT 22°C AMBIENT, USING NSF/ANSI 456 COMPLIANT VALIDATION BALLAST PROBES, EMPTY CHAMBER, DURING STABILIZED STEADY STATE OPERATION AND A DAQ SAMPLING RATE OF ONE MEASUREMENT EVERY 10 SECONDS
  1. UNIFORMITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE ACROSS ALL PROBES AT ANY POINT IN TIME OVER THE TESTING PERIOD
  2. STABILITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE EXPERIENCED BY ANY SINGLE PROBE OVER THE TESTING PERIOD
  3. TEMPERATURE PERFORMANCE FOR ALL LOADED AND UNLOADED DOOR OPENING PROTOCOLS, ALL ALARM, CONTROLLER AND PROBE REQUIREMENTS AS DEFINED IN THE NSF/ANSI 456 STANDARD FOR VACCINE STORAGE
  4. DATA PER ENERGY STAR TEST RESULTS OR EQUIVALENT TESTING AND CALCULATION.  HEAT REJECTION BASED ON DAILY AVERAGES, NOT CONTINUOUS OPERATION.  PERFORMANCE EXCEEDS ENERGY STAR REQUIREMENTS.
 CONTROLLER TECHNOLOGY PARAMETRIC, MICROPROCESSOR, LED DISPLAY   WITH 0.1°C RESOLUTION
 TEMPERATURE SETPOINT RANGE 1°C TO 10°C (SETPOINT MUST REMAIN   UNALTERED FROM THE FACTORY SETTING TO   REMAIN COMPLIANT WITH NSF/ANSI   PERFORMANCE REQUIREMENTS)
 DISPLAY PROBE CALIBRATED, STAINLESS STEEL
 EXTERNAL ALARM CONNECTION STATE SWITCHING REMOTE ALARM CONTACTS
 ALARMS VISUAL AND AUDIBLE INDICATORS
 HIGH / LOW TEMPERATURE, COMPLIANT WITH   ALARM REQUIREMENTS DEFINED IN THE   NSF/ANSI STANDARD FOR VACCINE STORAGE
 SIMULATOR BALLAST 20 ML BOTTLE, GLASS BEAD THERMAL MEDIA

Glass Door Freestanding Vaccine Refrigerator


These premier undercounter refrigerators are designed in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. Refrigerators protect pharmaceuticals at optimal temperatures, preventing waste and allowing for peak delivery.

These glass door freestanding refrigerators utilize microprocessor controllers and feature temperature alarms, remote alarm contacts, LED interior lighting, and probe access ports with included probes. American Biotech Supply Vaccine Storage Refrigerators utilize HFC-free refrigerant for environmental health and energy efficiency.
The acceptable temperature ranges for vaccine storage are 36°F to 46°F (or 2°C and 8°C) in the refrigerator and -58°F and 5°F (or -50°C and -15°C) in freezers.
*-one or more of these certifications may apply to this unit.

REFRIGERATION SYSTEM

 COMPRESSOR HERMETIC, HIGH PERFORMANCE
 REFRIGERANT EPA SNAP COMPLIANT, R600A,   ISOBUTANE
 CONDENSER TUBE AND GRID CONSTRUCTION,   FANLESS
 EVAPORATOR PLATE WALL
 DEFROST CYCLE OPTIMIZED, ZERO ENERGY
 
PERFORMANCE
 UNIFORMITY¹ (CABINET AIR) +/- 1.4°C
 STABILITY² (CABINET AIR) +/- 1.3°C
 MAXIMUM TEMPERATURE   VARIATION (CABINET AIR) +/- 1.7°C
 TEMPERATURE RISE AFTER AN   AFTER 8 SEC DOOR OPENINGS TEMPERATURE DID NOT EXCEED 6.4°C   AT ANY PROBE FOR ALL REQUIRED   NSF/ANSI 456 TESTING PROTOCOLS
 RECOVERY AFTER 3 MIN DOOR   OPENING ALL PROBES RECOVER TO UNDER 8°C   WITHIN 6 MIN.
 ENERGY CONSUMPTION 1.15 KWH/DAY
 AVERAGE HEAT REJECTION 1.67 KWH/DAY (224 BTU/H)
 NOISE PRESSURE LEVEL (DBA) 41 OR LESS INSTALLED
 PULL DOWN TIME TO NOMINAL   OPERATING TEMP 36 MIN


PERFORMANCE DATA ACQUIRED AT 22°C AMBIENT, USING NSF/ANSI 456 COMPLIANT VALIDATION BALLAST PROBES, EMPTY CHAMBER, DURING STABILIZED STEADY STATE OPERATION AND A DAQ SAMPLING RATE OF ONE MEASUREMENT EVERY 10 SECONDS
  • UNIFORMITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE ACROSS ALL PROBES AT ANY POINT IN TIME OVER THE TESTING PERIOD
  • STABILITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE EXPERIENCED BY ANY SINGLE PROBE OVER THE TESTING PERIOD
  • TEMPERATURE PERFORMANCE FOR ALL LOADED AND UNLOADED DOOR OPENING PROTOCOLS, ALL ALARM, CONTROLLER AND PROBE REQUIREMENTS AS DEFINED IN THE NSF/ANSI 456 STANDARD FOR VACCINE STORAGE
  • DATA PER ENERGY STAR TEST RESULTS OR EQUIVALENT TESTING AND CALCULATION.  HEAT REJECTION BASED ON DAILY AVERAGES, NOT CONTINUOUS OPERATION.  PERFORMANCE EXCEEDS ENERGY STAR REQUIREMENTS.

 CONTROLLER TECHNOLOGY PARAMETRIC, MICROPROCESSOR, LED DISPLAY   WITH 0.1°C RESOLUTION
 TEMPERATURE SETPOINT RANGE 1°C TO 10°C (SETPOINT MUST REMAIN   UNALTERED FROM THE FACTORY SETTING TO   REMAIN COMPLIANT WITH NSF/ANSI   PERFORMANCE REQUIREMENTS)
 DISPLAY PROBE CALIBRATED, STAINLESS STEEL
 EXTERNAL ALARM CONNECTION STATE SWITCHING REMOTE ALARM CONTACTS
 ALARMS VISUAL AND AUDIBLE INDICATORS
 HIGH / LOW TEMPERATURE, COMPLIANT WITH   ALARM REQUIREMENTS DEFINED IN THE   NSF/ANSI STANDARD FOR VACCINE STORAGE
 SIMULATOR BALLAST 20 ML BOTTLE, GLASS BEAD THERMAL MEDIA

Solid Door Freestanding Vaccine Refrigerator Certified to NSF/ANSI 456


These premier undercounter refrigerators are designed in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. Refrigerators protect pharmaceuticals at optimal temperatures, preventing waste and allowing for peak delivery.
These solid door freestanding refrigerators utilize microprocessor controllers and feature temperature alarms, remote alarm contacts, LED interior lighting, and probe access ports with included probes. American Biotech Supply Vaccine Storage Refrigerators utilize HFC-free refrigerant for environmental health and energy efficiency.
\The acceptable temperature ranges for vaccine storage are 36°F to 46°F (or 2°C and 8°C) in the refrigerator and -58°F and 5°F (or -50°C and -15°C) in freezers.

REFRIGERATION SYSTEM

 COMPRESSOR HERMETIC, HIGH PERFORMANCE
 REFRIGERANT EPA SNAP COMPLIANT, R600A, PROPANE
 CONDENSER TUBE AND GRID CONSTRUCTION,   FANLESS
 EVAPORATOR PLATE WALL
 DEFROST CYCLE OPTIMIZED, ZERO ENERGY
 
PERFORMANCE
 UNIFORMITY¹ (CABINET AIR) +/- 1.4°C
 STABILITY² (CABINET AIR) +/- 1.3°C
 MAXIMUM TEMPERATURE   VARIATION (CABINET AIR) +/- 1.7°C
 TEMPERATURE RISE AFTER AN   AFTER 8 SEC DOOR OPENINGS TEMPERATURE DID NOT EXCEED 6.5°C   AT ANY PROBE FOR ALL REQUIRED   NSF/ANSI 456 TESTING PROTOCOLS
 RECOVERY AFTER 3 MIN DOOR   OPENING ALL PROBES RECOVER TO UNDER 8°C   WITHIN 6 MIN.
 ENERGY CONSUMPTION 1.15 KWH/DAY
 AVERAGE HEAT REJECTION 1.67 KWH/DAY (315 BTU/H)
 NOISE PRESSURE LEVEL (DBA) 41 OR LESS INSTALLED
 PULL DOWN TIME TO NOMINAL   OPERATING TEMP 42 MIN


PERFORMANCE DATA ACQUIRED AT 22°C AMBIENT, USING NSF/ANSI 456 COMPLIANT VALIDATION BALLAST PROBES, EMPTY CHAMBER, DURING STABILIZED STEADY STATE OPERATION AND A DAQ SAMPLING RATE OF ONE MEASUREMENT EVERY 10 SECONDS
  • UNIFORMITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE ACROSS ALL PROBES AT ANY POINT IN TIME OVER THE TESTING PERIOD
  • STABILITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE EXPERIENCED BY ANY SINGLE PROBE OVER THE TESTING PERIOD
  • TEMPERATURE PERFORMANCE FOR ALL LOADED AND UNLOADED DOOR OPENING PROTOCOLS, ALL ALARM, CONTROLLER AND PROBE REQUIREMENTS AS DEFINED IN THE NSF/ANSI 456 STANDARD FOR VACCINE STORAGE
  • DATA PER ENERGY STAR TEST RESULTS OR EQUIVALENT TESTING AND CALCULATION.  HEAT REJECTION BASED ON DAILY AVERAGES, NOT CONTINUOUS OPERATION.  PERFORMANCE EXCEEDS ENERGY STAR REQUIREMENTS.

 CONTROLLER TECHNOLOGY PARAMETRIC, MICROPROCESSOR, LED DISPLAY   WITH 0.1°C RESOLUTION
 TEMPERATURE SETPOINT RANGE 1°C TO 10°C (SETPOINT MUST REMAIN   UNALTERED FROM THE FACTORY SETTING TO   REMAIN COMPLIANT WITH NSF/ANSI   PERFORMANCE REQUIREMENTS)
 DISPLAY PROBE CALIBRATED, STAINLESS STEEL
 EXTERNAL ALARM CONNECTION STATE SWITCHING REMOTE ALARM CONTACTS
 ALARMS VISUAL AND AUDIBLE INDICATORS
 HIGH / LOW TEMPERATURE, COMPLIANT WITH   ALARM REQUIREMENTS DEFINED IN THE   NSF/ANSI STANDARD FOR VACCINE STORAGE
 SIMULATOR BALLAST 20 ML BOTTLE, GLASS BEAD THERMAL MEDIA

Stainless Steel Built-In Vaccine Refrigerator


These premier built-in undercounter refrigerators are designed in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. Refrigerators protect pharmaceuticals at optimal temperatures, preventing waste and allowing for peak delivery.
The stainless steel Refrigerators utilize microprocessor controllers and feature temperature alarms, remote alarm contacts, and probe access ports with included probes. American Biotech Supply Vaccine Storage Refrigerators utilize HFC-free refrigerant for environmental health and energy efficiency.

REFRIGERATION SYSTEM

 COMPRESSOR HERMETIC, HIGH PERFORMANCE
 REFRIGERANT EPA SNAP COMPLIANT, R600A,   ISOBUTANE
 CONDENSER HYBRID FIN AND TUBE WITH LOW NOISE   FAN
 EVAPORATOR PLATE WALL
 DEFROST CYCLE OPTIMIZED, ZERO ENERGY
 
PERFORMANCE
 UNIFORMITY¹ (CABINET AIR) +/- 0.8°C
 STABILITY² (CABINET AIR) +/- 1.2°C
 MAXIMUM TEMPERATURE   VARIATION (CABINET AIR) +/- 1.4°C
 TEMPERATURE RISE AFTER AN   AFTER 8 SEC DOOR OPENINGS TEMPERATURE DID NOT EXCEED 6.4°C   AT ANY PROBE FOR ALL REQUIRED   NSF/ANSI 456 TESTING PROTOCOLS
 RECOVERY AFTER 3 MIN DOOR   OPENING ALL PROBES RECOVER TO UNDER 8°C   WITHIN 4.8 MIN.
 ENERGY CONSUMPTION 1.15 KWH/DAY
 AVERAGE HEAT REJECTION 1.57 KWH/DAY (224 BTU/H)
 NOISE PRESSURE LEVEL (DBA) 43 OR LESS INSTALLED
 PULL DOWN TIME TO NOMINAL   OPERATING TEMP 35 MIN


PERFORMANCE DATA ACQUIRED AT 22°C AMBIENT, USING NSF/ANSI 456 COMPLIANT VALIDATION BALLAST PROBES, EMPTY CHAMBER, DURING STABILIZED STEADY STATE OPERATION AND A DAQ SAMPLING RATE OF ONE MEASUREMENT EVERY 10 SECONDS
  • UNIFORMITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE ACROSS ALL PROBES AT ANY POINT IN TIME OVER THE TESTING PERIOD
  • STABILITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE EXPERIENCED BY ANY SINGLE PROBE OVER THE TESTING PERIOD
  • TEMPERATURE PERFORMANCE FOR ALL LOADED AND UNLOADED DOOR OPENING PROTOCOLS, ALL ALARM, CONTROLLER AND PROBE REQUIREMENTS AS DEFINED IN THE NSF/ANSI 456 STANDARD FOR VACCINE STORAGE
  • DATA PER ENERGY STAR TEST RESULTS OR EQUIVALENT TESTING AND CALCULATION.  HEAT REJECTION BASED ON DAILY AVERAGES, NOT CONTINUOUS OPERATION.  PERFORMANCE EXCEEDS ENERGY STAR REQUIREMENTS.

 CONTROLLER TECHNOLOGY PARAMETRIC, MICROPROCESSOR, LED DISPLAY   WITH 0.1°C RESOLUTION
 TEMPERATURE SETPOINT RANGE 1°C TO 10°C (SETPOINT MUST REMAIN   UNALTERED FROM THE FACTORY SETTING TO   REMAIN COMPLIANT WITH NSF/ANSI   PERFORMANCE REQUIREMENTS)
 DISPLAY PROBE CALIBRATED, STAINLESS STEEL
 EXTERNAL ALARM CONNECTION STATE SWITCHING REMOTE ALARM CONTACTS
 ALARMS VISUAL AND AUDIBLE INDICATORS
 HIGH / LOW TEMPERATURE, COMPLIANT WITH   ALARM REQUIREMENTS DEFINED IN THE   NSF/ANSI STANDARD FOR VACCINE STORAGE
 SIMULATOR BALLAST 20 ML BOTTLE, GLASS BEAD THERMAL MEDIA

 

Glass Door Built-in Vaccine Refrigerator


These premier built-in undercounter refrigerators are designed in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. Refrigerators protect pharmaceuticals at optimal temperatures, preventing waste and allowing for peak delivery.
The glass door refrigerators utilize microprocessor controllers and feature temperature alarms, remote alarm contacts, LED interior lighting, and probe access ports with included probes.

REFRIGERATION SYSTEM

 COMPRESSOR HERMETIC, HIGH PERFORMANCE
 REFRIGERANT EPA SNAP COMPLIANT, R600A,   ISOBUTANE
 CONDENSER HYBRID FIN AND TUBE WITH LOW NOISE   FAN
 EVAPORATOR PLATE WALL
 DEFROST CYCLE OPTIMIZED, ZERO ENERGY
 
PERFORMANCE
 UNIFORMITY¹ (CABINET AIR) +/- 0.8°C
 STABILITY² (CABINET AIR) +/- 1.2°C
 MAXIMUM TEMPERATURE   VARIATION (CABINET AIR) +/- 1.4°C
 TEMPERATURE RISE AFTER AN   AFTER 8 SEC DOOR OPENINGS TEMPERATURE DID NOT EXCEED 6.4°C   AT ANY PROBE FOR ALL REQUIRED   NSF/ANSI 456 TESTING PROTOCOLS
 RECOVERY AFTER 3 MIN DOOR   OPENING ALL PROBES RECOVER TO UNDER 8°C   WITHIN 4.8 MIN.
 ENERGY CONSUMPTION 1.15 KWH/DAY
 AVERAGE HEAT REJECTION 1.57 KWH/DAY (224 BTU/H)
 NOISE PRESSURE LEVEL (DBA) 43 OR LESS INSTALLED
 PULL DOWN TIME TO NOMINAL   OPERATING TEMP 35 MIN


PERFORMANCE DATA ACQUIRED AT 22°C AMBIENT, USING NSF/ANSI 456 COMPLIANT VALIDATION BALLAST PROBES, EMPTY CHAMBER, DURING STABILIZED STEADY STATE OPERATION AND A DAQ SAMPLING RATE OF ONE MEASUREMENT EVERY 10 SECONDS
  • UNIFORMITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE ACROSS ALL PROBES AT ANY POINT IN TIME OVER THE TESTING PERIOD
  • STABILITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE EXPERIENCED BY ANY SINGLE PROBE OVER THE TESTING PERIOD
  • TEMPERATURE PERFORMANCE FOR ALL LOADED AND UNLOADED DOOR OPENING PROTOCOLS, ALL ALARM, CONTROLLER AND PROBE REQUIREMENTS AS DEFINED IN THE NSF/ANSI 456 STANDARD FOR VACCINE STORAGE
  • DATA PER ENERGY STAR TEST RESULTS OR EQUIVALENT TESTING AND CALCULATION.  HEAT REJECTION BASED ON DAILY AVERAGES, NOT CONTINUOUS OPERATION.  PERFORMANCE EXCEEDS ENERGY STAR REQUIREMENTS.

 CONTROLLER TECHNOLOGY PARAMETRIC, MICROPROCESSOR, LED DISPLAY   WITH 0.1°C RESOLUTION
 TEMPERATURE SETPOINT RANGE 1°C TO 10°C (SETPOINT MUST REMAIN   UNALTERED FROM THE FACTORY SETTING TO   REMAIN COMPLIANT WITH NSF/ANSI   PERFORMANCE REQUIREMENTS)
 DISPLAY PROBE CALIBRATED, STAINLESS STEEL
 EXTERNAL ALARM CONNECTION STATE SWITCHING REMOTE ALARM CONTACTS
 ALARMS VISUAL AND AUDIBLE INDICATORS
 HIGH / LOW TEMPERATURE, COMPLIANT WITH   ALARM REQUIREMENTS DEFINED IN THE   NSF/ANSI STANDARD FOR VACCINE STORAGE
 SIMULATOR BALLAST 20 ML BOTTLE, GLASS BEAD THERMAL MEDIA

Built-In Vaccine Refrigerator NSF/ANSI 456 Certified


These premier built-in undercounter refrigerators are designed in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. Refrigerators protect pharmaceuticals at optimal temperatures, preventing waste and allowing for peak delivery.
The solid door refrigerators utilize microprocessor controllers and feature temperature alarms, remote alarm contacts, and probe access ports with included probes. American Biotech Supply Vaccine Storage Refrigerators utilize HFC-free refrigerant for environmental health and energy efficiency.

REFRIGERATION SYSTEM

 COMPRESSOR HERMETIC, HIGH PERFORMANCE
 REFRIGERANT EPA SNAP COMPLIANT, R600A,   ISOBUTANE
 CONDENSER HYBRID FIN AND TUBE WITH LOW NOISE   FAN
 EVAPORATOR PLATE WALL
 DEFROST CYCLE OPTIMIZED, ZERO ENERGY
 
PERFORMANCE
 UNIFORMITY¹ (CABINET AIR) +/- 0.8°C
 STABILITY² (CABINET AIR) +/- 1.2°C
 MAXIMUM TEMPERATURE   VARIATION (CABINET AIR) +/- 1.4°C
 TEMPERATURE RISE AFTER AN   AFTER 8 SEC DOOR OPENINGS TEMPERATURE DID NOT EXCEED 6.4°C   AT ANY PROBE FOR ALL REQUIRED   NSF/ANSI 456 TESTING PROTOCOLS
 RECOVERY AFTER 3 MIN DOOR   OPENING ALL PROBES RECOVER TO UNDER 8°C   WITHIN 4.8 MIN.
 ENERGY CONSUMPTION 1.15 KWH/DAY
 AVERAGE HEAT REJECTION 1.57 KWH/DAY (224 BTU/H)
 NOISE PRESSURE LEVEL (DBA) 43 OR LESS INSTALLED
 PULL DOWN TIME TO NOMINAL   OPERATING TEMP 35 MIN


PERFORMANCE DATA ACQUIRED AT 22°C AMBIENT, USING NSF/ANSI 456 COMPLIANT VALIDATION BALLAST PROBES, EMPTY CHAMBER, DURING STABILIZED STEADY STATE OPERATION AND A DAQ SAMPLING RATE OF ONE MEASUREMENT EVERY 10 SECONDS
  • UNIFORMITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE ACROSS ALL PROBES AT ANY POINT IN TIME OVER THE TESTING PERIOD
  • STABILITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE EXPERIENCED BY ANY SINGLE PROBE OVER THE TESTING PERIOD
  • TEMPERATURE PERFORMANCE FOR ALL LOADED AND UNLOADED DOOR OPENING PROTOCOLS, ALL ALARM, CONTROLLER AND PROBE REQUIREMENTS AS DEFINED IN THE NSF/ANSI 456 STANDARD FOR VACCINE STORAGE
  • DATA PER ENERGY STAR TEST RESULTS OR EQUIVALENT TESTING AND CALCULATION.  HEAT REJECTION BASED ON DAILY AVERAGES, NOT CONTINUOUS OPERATION.  PERFORMANCE EXCEEDS ENERGY STAR REQUIREMENTS.

 CONTROLLER TECHNOLOGY PARAMETRIC, MICROPROCESSOR, LED DISPLAY   WITH 0.1°C RESOLUTION
 TEMPERATURE SETPOINT RANGE 1°C TO 10°C (SETPOINT MUST REMAIN   UNALTERED FROM THE FACTORY SETTING TO   REMAIN COMPLIANT WITH NSF/ANSI   PERFORMANCE REQUIREMENTS)
 DISPLAY PROBE CALIBRATED, STAINLESS STEEL
 EXTERNAL ALARM CONNECTION STATE SWITCHING REMOTE ALARM CONTACTS
 ALARMS VISUAL AND AUDIBLE INDICATORS
 HIGH / LOW TEMPERATURE, COMPLIANT WITH   ALARM REQUIREMENTS DEFINED IN THE   NSF/ANSI STANDARD FOR VACCINE STORAGE
 SIMULATOR BALLAST 20 ML BOTTLE, GLASS BEAD THERMAL MEDIA