S2K Commerce - Products Dropdown
ActionsS2K Commerce - Shopping Cart
ActionsS2K Commerce - Order Entry
ActionsCO2 incubator w/ hot air sterilization
CO2 Incubator with Hot Air Sterilization
The new C 170 CO2 incubator has been designed for standard applications in the cell culture laboratory. It provides a reproducible control for the cell well-being parameters while also ensuring the best possible levels of contamination safety in the incubator thanks to its straightforward design. When stacked, it offers 11.8 cubic feet of incubation space and excellent usability.
BENEFITS:
- Best Uniformity in Class
- Increased Volume of 5.9 cubic feet
- Hot Air Sterilization Cycle at 180°C minimizing the risk of contamination
- Maximum Saturation Without Condensation
- Stable pH-values thanks to drift-free CO2 IR sensor technology
- Easy cleaning thanks to seamless, deep-drawn inner chamber
MAIN FEATURES:
- Temperature range: room temperature plus 7°C to 50°C
- Humidity range: up to 95 % RH
- Auto-sterilization with hot air at 180°C
- Humidification system with condensation protection
- CO2 gas-mixing jet with Venturi effect
- CO2 sensor with infrared technology
- Display with 3-digit, 7-segment LEDs
- Door hinged right or left
- Tightly-sealed inner door made of tempered safety glass
- Seamless, deep-drawn inner stainless steel chamber
- 3 perforated stainless steel shelves
- Units are stackable with stacking adapter
- Troubleshooting system with visual and audible alarms
- Zero-voltage alarm contact
TECHNICAL DATA:
Description |
9040-0113* |
9040-0116* |
Performance Data Temperature |
||
Temperature range 7°C above ambient temperature to [°C] |
50 |
50 |
Temperature variation at 37°C [± K] |
0.3 |
0.3 |
Temperature fluctuation [± K] |
0.1 |
0.1 |
Recovery time after 30 seconds door open at 5 Vol.-% CO2 [min] |
5 |
5 |
Recovery time after 30 seconds door open at 37 °C [min] |
6 |
6 |
Performance Data Climate |
||
Humidity range [% RH] |
90â¦95 |
90â¦95 |
Gas Data - CO2-Data |
||
CO2 range [Vol.-% CO2] |
0â¦20 |
0â¦20 |
CO2 measurement |
IR |
IR |
Electrical data |
||
Rated Voltage [V] |
120 |
120 |
Power frequency [Hz] |
60 |
60 |
Nominal power [kW] |
1.1 |
1.1 |
Unit fuse [A] |
10 |
10 |
Phase (Nominal voltage) |
1~ |
1~ |
Measures - Outer dimensions |
||
Width net [mm/in] |
740 /29.16 |
740 /29.16 |
Height net [mm/in] |
870/34.28 |
870/34.28 |
Wall clearance back [mm/in] |
100/3.94 |
100/3.94 |
Wall clearance sidewise [mm/in] |
50/1.97 |
50/1.97 |
Doors |
||
Inner doors |
1 |
1 |
Unit doors |
1 |
1 |
Internal Dimensions |
||
Width [mm/in] |
560/22.06 |
560/22.06 |
Height [mm/in] |
600/23.64 |
600/23.64 |
Depth [mm/in] |
500/19.7 |
500/19.7 |
Measures |
||
Interior volume [L/cu.ft.] |
168/5.9 |
168/5.9 |
Net weight of the unit (empty) [kg/lbs] |
110/243 |
110/243 |
Load per rack [kg/lbs] |
10/22 |
10/22 |
Environment-specific data |
||
Energy consumption at 37°C |
90 Wh/h |
90 Wh/h |
Fixtures |
||
Number of shelves (standard/ maximum) |
3/8 |
3/8 |
* All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10%. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. All indications are average values, typical for units produced in series.