S2K Commerce - Products Dropdown
S2K Commerce - Shopping Cart
S2K Commerce - Order Entry
Solid Door Built-In Vaccine Refrigerator
REFRIGERATION SYSTEM
| COMPRESSOR | HERMETIC, HIGH PERFORMANCE |
| REFRIGERANT | EPA SNAP COMPLIANT, R600A, ISOBUTANE |
| CONDENSER | HYBRID FIN AND TUBE WITH LOW NOISE FAN |
| EVAPORATOR | PLATE WALL |
| DEFROST | CYCLE OPTIMIZED, ZERO ENERGY |
| UNIFORMITY¹ (CABINET AIR) | +/- 0.8°C |
| STABILITY² (CABINET AIR) | +/- 1.2°C |
| MAXIMUM TEMPERATURE VARIATION (CABINET AIR) | +/- 1.4°C |
| TEMPERATURE RISE AFTER AN AFTER 8 SEC DOOR OPENINGS | TEMPERATURE DID NOT EXCEED 6.4°C AT ANY PROBE FOR ALL REQUIRED NSF/ANSI 456 TESTING PROTOCOLS |
| RECOVERY AFTER 3 MIN DOOR OPENING | ALL PROBES RECOVER TO UNDER 8°C WITHIN 4.8 MIN. |
| ENERGY CONSUMPTION | 1.15 KWH/DAY |
| AVERAGE HEAT REJECTION | 1.57 KWH/DAY (224 BTU/H) |
| NOISE PRESSURE LEVEL (DBA) | 43 OR LESS INSTALLED |
| PULL DOWN TIME TO NOMINAL OPERATING TEMP | 35 MIN |
- UNIFORMITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE ACROSS ALL PROBES AT ANY POINT IN TIME OVER THE TESTING PERIOD
- STABILITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE EXPERIENCED BY ANY SINGLE PROBE OVER THE TESTING PERIOD
- TEMPERATURE PERFORMANCE FOR ALL LOADED AND UNLOADED DOOR OPENING PROTOCOLS, ALL ALARM, CONTROLLER AND PROBE REQUIREMENTS AS DEFINED IN THE NSF/ANSI 456 STANDARD FOR VACCINE STORAGE
- DATA PER ENERGY STAR TEST RESULTS OR EQUIVALENT TESTING AND CALCULATION. HEAT REJECTION BASED ON DAILY AVERAGES, NOT CONTINUOUS OPERATION. PERFORMANCE EXCEEDS ENERGY STAR REQUIREMENTS.
| CONTROLLER TECHNOLOGY | PARAMETRIC, MICROPROCESSOR, LED DISPLAY WITH 0.1°C RESOLUTION |
| TEMPERATURE SETPOINT RANGE | 1°C TO 10°C (SETPOINT MUST REMAIN UNALTERED FROM THE FACTORY SETTING TO REMAIN COMPLIANT WITH NSF/ANSI PERFORMANCE REQUIREMENTS) |
| DISPLAY PROBE | CALIBRATED, STAINLESS STEEL |
| EXTERNAL ALARM CONNECTION | STATE SWITCHING REMOTE ALARM CONTACTS |
| ALARMS | VISUAL AND AUDIBLE INDICATORS |
| HIGH / LOW TEMPERATURE, COMPLIANT WITH ALARM REQUIREMENTS DEFINED IN THE NSF/ANSI STANDARD FOR VACCINE STORAGE | |
| SIMULATOR BALLAST | 20 ML BOTTLE, GLASS BEAD THERMAL MEDIA |
Glass Door Freestanding Vaccine Refrigerator
| COMPRESSOR | HERMETIC, HIGH PERFORMANCE |
| REFRIGERANT | EPA SNAP COMPLIANT, R600A, ISOBUTANE |
| CONDENSER | TUBE AND GRID CONSTRUCTION, FANLESS |
| EVAPORATOR | PLATE WALL |
| DEFROST | CYCLE OPTIMIZED, ZERO ENERGY |
| UNIFORMITY¹ (CABINET AIR) | +/- 1.4°C |
| STABILITY² (CABINET AIR) | +/- 1.3°C |
| MAXIMUM TEMPERATURE VARIATION (CABINET AIR) | +/- 1.7°C |
| TEMPERATURE RISE AFTER AN AFTER 8 SEC DOOR OPENINGS | TEMPERATURE DID NOT EXCEED 6.4°C AT ANY PROBE FOR ALL REQUIRED NSF/ANSI 456 TESTING PROTOCOLS |
| RECOVERY AFTER 3 MIN DOOR OPENING | ALL PROBES RECOVER TO UNDER 8°C WITHIN 6 MIN. |
| ENERGY CONSUMPTION | 1.15 KWH/DAY |
| AVERAGE HEAT REJECTION | 1.67 KWH/DAY (224 BTU/H) |
| NOISE PRESSURE LEVEL (DBA) | 41 OR LESS INSTALLED |
| PULL DOWN TIME TO NOMINAL OPERATING TEMP | 36 MIN |
PERFORMANCE DATA ACQUIRED AT 22°C AMBIENT, USING NSF/ANSI 456 COMPLIANT VALIDATION BALLAST PROBES, EMPTY CHAMBER, DURING STABILIZED STEADY STATE OPERATION AND A DAQ SAMPLING RATE OF ONE MEASUREMENT EVERY 10 SECONDS
- UNIFORMITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE ACROSS ALL PROBES AT ANY POINT IN TIME OVER THE TESTING PERIOD
- STABILITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE EXPERIENCED BY ANY SINGLE PROBE OVER THE TESTING PERIOD
- TEMPERATURE PERFORMANCE FOR ALL LOADED AND UNLOADED DOOR OPENING PROTOCOLS, ALL ALARM, CONTROLLER AND PROBE REQUIREMENTS AS DEFINED IN THE NSF/ANSI 456 STANDARD FOR VACCINE STORAGE
- DATA PER ENERGY STAR TEST RESULTS OR EQUIVALENT TESTING AND CALCULATION. HEAT REJECTION BASED ON DAILY AVERAGES, NOT CONTINUOUS OPERATION. PERFORMANCE EXCEEDS ENERGY STAR REQUIREMENTS.
| CONTROLLER TECHNOLOGY | PARAMETRIC, MICROPROCESSOR, LED DISPLAY WITH 0.1°C RESOLUTION |
| TEMPERATURE SETPOINT RANGE | 1°C TO 10°C (SETPOINT MUST REMAIN UNALTERED FROM THE FACTORY SETTING TO REMAIN COMPLIANT WITH NSF/ANSI PERFORMANCE REQUIREMENTS) |
| DISPLAY PROBE | CALIBRATED, STAINLESS STEEL |
| EXTERNAL ALARM CONNECTION | STATE SWITCHING REMOTE ALARM CONTACTS |
| ALARMS | VISUAL AND AUDIBLE INDICATORS |
| HIGH / LOW TEMPERATURE, COMPLIANT WITH ALARM REQUIREMENTS DEFINED IN THE NSF/ANSI STANDARD FOR VACCINE STORAGE | |
| SIMULATOR BALLAST | 20 ML BOTTLE, GLASS BEAD THERMAL MEDIA |
Solid Door Freestanding Vaccine Refrigerator Certified to NSF/ANSI 456
| COMPRESSOR | HERMETIC, HIGH PERFORMANCE |
| REFRIGERANT | EPA SNAP COMPLIANT, R600A, PROPANE |
| CONDENSER | TUBE AND GRID CONSTRUCTION, FANLESS |
| EVAPORATOR | PLATE WALL |
| DEFROST | CYCLE OPTIMIZED, ZERO ENERGY |
| UNIFORMITY¹ (CABINET AIR) | +/- 1.4°C |
| STABILITY² (CABINET AIR) | +/- 1.3°C |
| MAXIMUM TEMPERATURE VARIATION (CABINET AIR) | +/- 1.7°C |
| TEMPERATURE RISE AFTER AN AFTER 8 SEC DOOR OPENINGS | TEMPERATURE DID NOT EXCEED 6.5°C AT ANY PROBE FOR ALL REQUIRED NSF/ANSI 456 TESTING PROTOCOLS |
| RECOVERY AFTER 3 MIN DOOR OPENING | ALL PROBES RECOVER TO UNDER 8°C WITHIN 6 MIN. |
| ENERGY CONSUMPTION | 1.15 KWH/DAY |
| AVERAGE HEAT REJECTION | 1.67 KWH/DAY (315 BTU/H) |
| NOISE PRESSURE LEVEL (DBA) | 41 OR LESS INSTALLED |
| PULL DOWN TIME TO NOMINAL OPERATING TEMP | 42 MIN |
PERFORMANCE DATA ACQUIRED AT 22°C AMBIENT, USING NSF/ANSI 456 COMPLIANT VALIDATION BALLAST PROBES, EMPTY CHAMBER, DURING STABILIZED STEADY STATE OPERATION AND A DAQ SAMPLING RATE OF ONE MEASUREMENT EVERY 10 SECONDS
- UNIFORMITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE ACROSS ALL PROBES AT ANY POINT IN TIME OVER THE TESTING PERIOD
- STABILITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE EXPERIENCED BY ANY SINGLE PROBE OVER THE TESTING PERIOD
- TEMPERATURE PERFORMANCE FOR ALL LOADED AND UNLOADED DOOR OPENING PROTOCOLS, ALL ALARM, CONTROLLER AND PROBE REQUIREMENTS AS DEFINED IN THE NSF/ANSI 456 STANDARD FOR VACCINE STORAGE
- DATA PER ENERGY STAR TEST RESULTS OR EQUIVALENT TESTING AND CALCULATION. HEAT REJECTION BASED ON DAILY AVERAGES, NOT CONTINUOUS OPERATION. PERFORMANCE EXCEEDS ENERGY STAR REQUIREMENTS.
| CONTROLLER TECHNOLOGY | PARAMETRIC, MICROPROCESSOR, LED DISPLAY WITH 0.1°C RESOLUTION |
| TEMPERATURE SETPOINT RANGE | 1°C TO 10°C (SETPOINT MUST REMAIN UNALTERED FROM THE FACTORY SETTING TO REMAIN COMPLIANT WITH NSF/ANSI PERFORMANCE REQUIREMENTS) |
| DISPLAY PROBE | CALIBRATED, STAINLESS STEEL |
| EXTERNAL ALARM CONNECTION | STATE SWITCHING REMOTE ALARM CONTACTS |
| ALARMS | VISUAL AND AUDIBLE INDICATORS |
| HIGH / LOW TEMPERATURE, COMPLIANT WITH ALARM REQUIREMENTS DEFINED IN THE NSF/ANSI STANDARD FOR VACCINE STORAGE | |
| SIMULATOR BALLAST | 20 ML BOTTLE, GLASS BEAD THERMAL MEDIA |
Stainless Steel Built-In Vaccine Refrigerator
| COMPRESSOR | HERMETIC, HIGH PERFORMANCE |
| REFRIGERANT | EPA SNAP COMPLIANT, R600A, ISOBUTANE |
| CONDENSER | HYBRID FIN AND TUBE WITH LOW NOISE FAN |
| EVAPORATOR | PLATE WALL |
| DEFROST | CYCLE OPTIMIZED, ZERO ENERGY |
| UNIFORMITY¹ (CABINET AIR) | +/- 0.8°C |
| STABILITY² (CABINET AIR) | +/- 1.2°C |
| MAXIMUM TEMPERATURE VARIATION (CABINET AIR) | +/- 1.4°C |
| TEMPERATURE RISE AFTER AN AFTER 8 SEC DOOR OPENINGS | TEMPERATURE DID NOT EXCEED 6.4°C AT ANY PROBE FOR ALL REQUIRED NSF/ANSI 456 TESTING PROTOCOLS |
| RECOVERY AFTER 3 MIN DOOR OPENING | ALL PROBES RECOVER TO UNDER 8°C WITHIN 4.8 MIN. |
| ENERGY CONSUMPTION | 1.15 KWH/DAY |
| AVERAGE HEAT REJECTION | 1.57 KWH/DAY (224 BTU/H) |
| NOISE PRESSURE LEVEL (DBA) | 43 OR LESS INSTALLED |
| PULL DOWN TIME TO NOMINAL OPERATING TEMP | 35 MIN |
PERFORMANCE DATA ACQUIRED AT 22°C AMBIENT, USING NSF/ANSI 456 COMPLIANT VALIDATION BALLAST PROBES, EMPTY CHAMBER, DURING STABILIZED STEADY STATE OPERATION AND A DAQ SAMPLING RATE OF ONE MEASUREMENT EVERY 10 SECONDS
- UNIFORMITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE ACROSS ALL PROBES AT ANY POINT IN TIME OVER THE TESTING PERIOD
- STABILITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE EXPERIENCED BY ANY SINGLE PROBE OVER THE TESTING PERIOD
- TEMPERATURE PERFORMANCE FOR ALL LOADED AND UNLOADED DOOR OPENING PROTOCOLS, ALL ALARM, CONTROLLER AND PROBE REQUIREMENTS AS DEFINED IN THE NSF/ANSI 456 STANDARD FOR VACCINE STORAGE
- DATA PER ENERGY STAR TEST RESULTS OR EQUIVALENT TESTING AND CALCULATION. HEAT REJECTION BASED ON DAILY AVERAGES, NOT CONTINUOUS OPERATION. PERFORMANCE EXCEEDS ENERGY STAR REQUIREMENTS.
| CONTROLLER TECHNOLOGY | PARAMETRIC, MICROPROCESSOR, LED DISPLAY WITH 0.1°C RESOLUTION |
| TEMPERATURE SETPOINT RANGE | 1°C TO 10°C (SETPOINT MUST REMAIN UNALTERED FROM THE FACTORY SETTING TO REMAIN COMPLIANT WITH NSF/ANSI PERFORMANCE REQUIREMENTS) |
| DISPLAY PROBE | CALIBRATED, STAINLESS STEEL |
| EXTERNAL ALARM CONNECTION | STATE SWITCHING REMOTE ALARM CONTACTS |
| ALARMS | VISUAL AND AUDIBLE INDICATORS |
| HIGH / LOW TEMPERATURE, COMPLIANT WITH ALARM REQUIREMENTS DEFINED IN THE NSF/ANSI STANDARD FOR VACCINE STORAGE | |
| SIMULATOR BALLAST | 20 ML BOTTLE, GLASS BEAD THERMAL MEDIA |
Glass Door Built-in Vaccine Refrigerator
| COMPRESSOR | HERMETIC, HIGH PERFORMANCE |
| REFRIGERANT | EPA SNAP COMPLIANT, R600A, ISOBUTANE |
| CONDENSER | HYBRID FIN AND TUBE WITH LOW NOISE FAN |
| EVAPORATOR | PLATE WALL |
| DEFROST | CYCLE OPTIMIZED, ZERO ENERGY |
| UNIFORMITY¹ (CABINET AIR) | +/- 0.8°C |
| STABILITY² (CABINET AIR) | +/- 1.2°C |
| MAXIMUM TEMPERATURE VARIATION (CABINET AIR) | +/- 1.4°C |
| TEMPERATURE RISE AFTER AN AFTER 8 SEC DOOR OPENINGS | TEMPERATURE DID NOT EXCEED 6.4°C AT ANY PROBE FOR ALL REQUIRED NSF/ANSI 456 TESTING PROTOCOLS |
| RECOVERY AFTER 3 MIN DOOR OPENING | ALL PROBES RECOVER TO UNDER 8°C WITHIN 4.8 MIN. |
| ENERGY CONSUMPTION | 1.15 KWH/DAY |
| AVERAGE HEAT REJECTION | 1.57 KWH/DAY (224 BTU/H) |
| NOISE PRESSURE LEVEL (DBA) | 43 OR LESS INSTALLED |
| PULL DOWN TIME TO NOMINAL OPERATING TEMP | 35 MIN |
PERFORMANCE DATA ACQUIRED AT 22°C AMBIENT, USING NSF/ANSI 456 COMPLIANT VALIDATION BALLAST PROBES, EMPTY CHAMBER, DURING STABILIZED STEADY STATE OPERATION AND A DAQ SAMPLING RATE OF ONE MEASUREMENT EVERY 10 SECONDS
- UNIFORMITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE ACROSS ALL PROBES AT ANY POINT IN TIME OVER THE TESTING PERIOD
- STABILITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE EXPERIENCED BY ANY SINGLE PROBE OVER THE TESTING PERIOD
- TEMPERATURE PERFORMANCE FOR ALL LOADED AND UNLOADED DOOR OPENING PROTOCOLS, ALL ALARM, CONTROLLER AND PROBE REQUIREMENTS AS DEFINED IN THE NSF/ANSI 456 STANDARD FOR VACCINE STORAGE
- DATA PER ENERGY STAR TEST RESULTS OR EQUIVALENT TESTING AND CALCULATION. HEAT REJECTION BASED ON DAILY AVERAGES, NOT CONTINUOUS OPERATION. PERFORMANCE EXCEEDS ENERGY STAR REQUIREMENTS.
| CONTROLLER TECHNOLOGY | PARAMETRIC, MICROPROCESSOR, LED DISPLAY WITH 0.1°C RESOLUTION |
| TEMPERATURE SETPOINT RANGE | 1°C TO 10°C (SETPOINT MUST REMAIN UNALTERED FROM THE FACTORY SETTING TO REMAIN COMPLIANT WITH NSF/ANSI PERFORMANCE REQUIREMENTS) |
| DISPLAY PROBE | CALIBRATED, STAINLESS STEEL |
| EXTERNAL ALARM CONNECTION | STATE SWITCHING REMOTE ALARM CONTACTS |
| ALARMS | VISUAL AND AUDIBLE INDICATORS |
| HIGH / LOW TEMPERATURE, COMPLIANT WITH ALARM REQUIREMENTS DEFINED IN THE NSF/ANSI STANDARD FOR VACCINE STORAGE | |
| SIMULATOR BALLAST | 20 ML BOTTLE, GLASS BEAD THERMAL MEDIA |
Built-In Vaccine Refrigerator NSF/ANSI 456 Certified
| COMPRESSOR | HERMETIC, HIGH PERFORMANCE |
| REFRIGERANT | EPA SNAP COMPLIANT, R600A, ISOBUTANE |
| CONDENSER | HYBRID FIN AND TUBE WITH LOW NOISE FAN |
| EVAPORATOR | PLATE WALL |
| DEFROST | CYCLE OPTIMIZED, ZERO ENERGY |
| UNIFORMITY¹ (CABINET AIR) | +/- 0.8°C |
| STABILITY² (CABINET AIR) | +/- 1.2°C |
| MAXIMUM TEMPERATURE VARIATION (CABINET AIR) | +/- 1.4°C |
| TEMPERATURE RISE AFTER AN AFTER 8 SEC DOOR OPENINGS | TEMPERATURE DID NOT EXCEED 6.4°C AT ANY PROBE FOR ALL REQUIRED NSF/ANSI 456 TESTING PROTOCOLS |
| RECOVERY AFTER 3 MIN DOOR OPENING | ALL PROBES RECOVER TO UNDER 8°C WITHIN 4.8 MIN. |
| ENERGY CONSUMPTION | 1.15 KWH/DAY |
| AVERAGE HEAT REJECTION | 1.57 KWH/DAY (224 BTU/H) |
| NOISE PRESSURE LEVEL (DBA) | 43 OR LESS INSTALLED |
| PULL DOWN TIME TO NOMINAL OPERATING TEMP | 35 MIN |
PERFORMANCE DATA ACQUIRED AT 22°C AMBIENT, USING NSF/ANSI 456 COMPLIANT VALIDATION BALLAST PROBES, EMPTY CHAMBER, DURING STABILIZED STEADY STATE OPERATION AND A DAQ SAMPLING RATE OF ONE MEASUREMENT EVERY 10 SECONDS
- UNIFORMITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE ACROSS ALL PROBES AT ANY POINT IN TIME OVER THE TESTING PERIOD
- STABILITY IS DEFINED AS THE MAXIMUM VARIANCE IN TEMPERATURE EXPERIENCED BY ANY SINGLE PROBE OVER THE TESTING PERIOD
- TEMPERATURE PERFORMANCE FOR ALL LOADED AND UNLOADED DOOR OPENING PROTOCOLS, ALL ALARM, CONTROLLER AND PROBE REQUIREMENTS AS DEFINED IN THE NSF/ANSI 456 STANDARD FOR VACCINE STORAGE
- DATA PER ENERGY STAR TEST RESULTS OR EQUIVALENT TESTING AND CALCULATION. HEAT REJECTION BASED ON DAILY AVERAGES, NOT CONTINUOUS OPERATION. PERFORMANCE EXCEEDS ENERGY STAR REQUIREMENTS.
| CONTROLLER TECHNOLOGY | PARAMETRIC, MICROPROCESSOR, LED DISPLAY WITH 0.1°C RESOLUTION |
| TEMPERATURE SETPOINT RANGE | 1°C TO 10°C (SETPOINT MUST REMAIN UNALTERED FROM THE FACTORY SETTING TO REMAIN COMPLIANT WITH NSF/ANSI PERFORMANCE REQUIREMENTS) |
| DISPLAY PROBE | CALIBRATED, STAINLESS STEEL |
| EXTERNAL ALARM CONNECTION | STATE SWITCHING REMOTE ALARM CONTACTS |
| ALARMS | VISUAL AND AUDIBLE INDICATORS |
| HIGH / LOW TEMPERATURE, COMPLIANT WITH ALARM REQUIREMENTS DEFINED IN THE NSF/ANSI STANDARD FOR VACCINE STORAGE | |
| SIMULATOR BALLAST | 20 ML BOTTLE, GLASS BEAD THERMAL MEDIA |
PERFORMANCE PHARMACY-VACCINE REFRIGERATORS
Designed and purpose-built for pharmacy, medication, and vaccination applications to support meeting CDC/VFC vaccine storage guidelines.
FEATURING:
- Built-in readout for current and high/low temperature
- Audible and visual high/low temperature alarms
- Optimized efficient performance for stable temperature
- Eco-friendly natural refrigerants for a greener footprint
ADVANCED TEMPERATURE CONTROL
- Intelligent microprocessor digital temperature controller
- Improved temperature accuracy with a buffered probe in glycol-filled bottle to simulate temperature of stored product
- Additional sensor reads interior air temperature for superior temperature control
- Adjustable operating control range from +2°C to +8°C (+35.6°F to +46.4°F)
- Digital display of the min/max temperature (updated in 15 minute increments)
- Temperature viewable in Celsius and Fahrenheit to the nearest tenth of a degree
- Password protection of control parameters beyond setpoint
- Optimized forced air cooling for excellent temperature stability and uniformity with rapid recovery after door openings and product loading
MULTIPLE ALARMS FOR SAFER MONITORING
- Audible/visual alarms for high/low temperature excursions
- Open door alarm sounds if the door is left ajar (default 4 minutes with selectable 1-99 minute range)
- Audible power failure and sensor failure alarms
- Mute function for alarms
- Remote alarm contacts in rear of the unit
FULLY FEATURED INTERIOR FOR CONVENIENT STORAGE
- Energy efficient LED lighting with internal on/off switch
- Adjustable plastic-coated wire shelves for improved circulation
- Multiple shelf positions spaced at ½ to allow flexible adjusting to support smaller items
- Adjustable cycle defrost operation
- probe port included to accept user-provided monitoring equipment
QUALITY DESIGN FOR RELIABLE PERFORMANCE
- Self-closing door offers added storage protection
- User-reversible door swing (self-closing function compatible with RHD position only)
- Environmentally-friendly natural hydrocarbon refrigerant for reduced global warming potential
- Meets EPA and SNAP requirements
- Convenient security with a top-mounted door lock (2 keys included)
- Magnetic gaskets to ensure a positive seal
- Pyxis®, Omnicell® and AcuDose RX® compatible
TECHNICAL SPECIFICIATIONS:
Model |
Capacity |
Height [in] |
Width [in] |
Depth [in] |
Door |
Cabinet |
Shelves |
Internal Fans |
ARS1PV |
1 Cu. Ft. |
21.5 |
17.5 |
19.5 |
White |
White |
3 |
1 |
ARG1PV |
1 Cu. Ft. |
21.5 |
17.5 |
20 |
Glass |
White |
3 |
1 |
ARS3PV |
3 Cu. Ft. |
33.75 |
18.5 |
19 |
White |
White |
4 |
1 |
ARG3PV |
3 Cu. Ft. |
33.75 |
18.5 |
19.5 |
Glass |
White |
4 |
1 |
ARS6PV |
6 Cu. Ft. |
32.5 |
23.5 |
24.5 |
White |
White |
4 |
1 |
ARG6PV |
6 Cu. Ft. |
32.5 |
23.5 |
25 |
Glass |
White |
4 |
1 |
ARS8PV |
8 Cu. Ft. |
50 |
23.38 |
24.5 |
White |
White |
7 |
2 |
ARG8PV |
8 Cu. Ft. |
50 |
23.38 |
25 |
Glass |
White |
7 |
2 |
ARS12PV |
12 Cu. Ft. |
61.75 |
23.38 |
24.5 |
White |
White |
7 |
2 |
ARG12PV |
12 Cu. Ft. |
61.75 |
23.38 |
25 |
Glass |
White |
7 |
2 |
ARS15PV |
15 Cu. Ft. |
72 |
23.5 |
24.5 |
White |
White |
7 |
2 |
ARG15PV |
15 Cu. Ft. |
72 |
23.5 |
25 |
Glass |
White |
7 |
2 |
TECHNICAL SPECIFICATIONS (ALL UNITS):
- Defrost: Cycle
- Thermostat: Intelligent microprocessor digital
- Temperature Sensors: 2 (air probe and buffered probe)
- Adjustable Operating Control Range: 2°C to +8°C (+35.6°F to +46.4°F)
- Preset: 5ºC/41ºF
- Temperature Variation: +/-1°C
- Shelving: White plastic coated wire
- Voltage: 115V/60Hz
- Cord: NEMA 5-15P
- Lock: Factory-installed key lock
- Refrigerant: Eco-friendly R600a
- Light: LED
- Access Port:
- Door: Reversible
- Fan: Internal
- Certifications: ETL, ETL-C
- Parts & Labor Warranty: 2 years
- Compressor Warranty: 5 years
INCLUDED
- DO NOT UNPLUG warning sticker
- DO NOT ADJUST TEMPERATURE warning sticker
- DO NOT STORE PERSONAL ITEMS warning sticker
OPTIONS
- NIST calibrated temperature readout with 2-year certificate of conformity
- Digital data logger (DDL) installed or as kit with 2-year certificate of conformity
PERFORMANCE PHARMACY-VACCINE REFRIGERATORS
Designed and purpose-built for pharmacy, medication, and vaccination applications to support meeting CDC/VFC vaccine storage guidelines.
FEATURING:
- Built-in readout for current and high/low temperature
- Audible and visual high/low temperature alarms
- Optimized efficient performance for stable temperature
- Eco-friendly natural refrigerants for a greener footprint
ADVANCED TEMPERATURE CONTROL
- Intelligent microprocessor digital temperature controller
- Improved temperature accuracy with a buffered probe in glycol-filled bottle to simulate temperature of stored product
- Additional sensor reads interior air temperature for superior temperature control
- Adjustable operating control range from +2°C to +8°C (+35.6°F to +46.4°F)
- Digital display of the min/max temperature (updated in 15 minute increments)
- Temperature viewable in Celsius and Fahrenheit to the nearest tenth of a degree
- Password protection of control parameters beyond setpoint
- Optimized forced air cooling for excellent temperature stability and uniformity with rapid recovery after door openings and product loading
MULTIPLE ALARMS FOR SAFER MONITORING
- Audible/visual alarms for high/low temperature excursions
- Open door alarm sounds if the door is left ajar (default 4 minutes with selectable 1-99 minute range)
- Audible power failure and sensor failure alarms
- Mute function for alarms
- Remote alarm contacts in rear of the unit
FULLY FEATURED INTERIOR FOR CONVENIENT STORAGE
- Energy efficient LED lighting with internal on/off switch
- Adjustable plastic-coated wire shelves for improved circulation
- Multiple shelf positions spaced at ½ to allow flexible adjusting to support smaller items
- Adjustable cycle defrost operation
- probe port included to accept user-provided monitoring equipment
QUALITY DESIGN FOR RELIABLE PERFORMANCE
- Self-closing door offers added storage protection
- User-reversible door swing (self-closing function compatible with RHD position only)
- Environmentally-friendly natural hydrocarbon refrigerant for reduced global warming potential
- Meets EPA and SNAP requirements
- Convenient security with a top-mounted door lock (2 keys included)
- Magnetic gaskets to ensure a positive seal
- Pyxis®, Omnicell® and AcuDose RX® compatible
TECHNICAL SPECIFICIATIONS:
Model |
Capacity |
Height [in] |
Width [in] |
Depth [in] |
Door |
Cabinet |
Shelves |
Internal Fans |
ARS1PV |
1 Cu. Ft. |
21.5 |
17.5 |
19.5 |
White |
White |
3 |
1 |
ARG1PV |
1 Cu. Ft. |
21.5 |
17.5 |
20 |
Glass |
White |
3 |
1 |
ARS3PV |
3 Cu. Ft. |
33.75 |
18.5 |
19 |
White |
White |
4 |
1 |
ARG3PV |
3 Cu. Ft. |
33.75 |
18.5 |
19.5 |
Glass |
White |
4 |
1 |
ARS6PV |
6 Cu. Ft. |
32.5 |
23.5 |
24.5 |
White |
White |
4 |
1 |
ARG6PV |
6 Cu. Ft. |
32.5 |
23.5 |
25 |
Glass |
White |
4 |
1 |
ARS8PV |
8 Cu. Ft. |
50 |
23.38 |
24.5 |
White |
White |
7 |
2 |
ARG8PV |
8 Cu. Ft. |
50 |
23.38 |
25 |
Glass |
White |
7 |
2 |
ARS12PV |
12 Cu. Ft. |
61.75 |
23.38 |
24.5 |
White |
White |
7 |
2 |
ARG12PV |
12 Cu. Ft. |
61.75 |
23.38 |
25 |
Glass |
White |
7 |
2 |
ARS15PV |
15 Cu. Ft. |
72 |
23.5 |
24.5 |
White |
White |
7 |
2 |
ARG15PV |
15 Cu. Ft. |
72 |
23.5 |
25 |
Glass |
White |
7 |
2 |
TECHNICAL SPECIFICATIONS (ALL UNITS):
- Defrost: Cycle
- Thermostat: Intelligent microprocessor digital
- Temperature Sensors: 2 (air probe and buffered probe)
- Adjustable Operating Control Range: 2°C to +8°C (+35.6°F to +46.4°F)
- Preset: 5ºC/41ºF
- Temperature Variation: +/-1°C
- Shelving: White plastic coated wire
- Voltage: 115V/60Hz
- Cord: NEMA 5-15P
- Lock: Factory-installed key lock
- Refrigerant: Eco-friendly R600a
- Light: LED
- Access Port:
- Door: Reversible
- Fan: Internal
- Certifications: ETL, ETL-C
- Parts & Labor Warranty: 2 years
- Compressor Warranty: 5 years
INCLUDED
- DO NOT UNPLUG warning sticker
- DO NOT ADJUST TEMPERATURE warning sticker
- DO NOT STORE PERSONAL ITEMS warning sticker
OPTIONS
- NIST calibrated temperature readout with 2-year certificate of conformity
- Digital data logger (DDL) installed or as kit with 2-year certificate of conformity
Premier Refrigerator With Built-In Pharmacy/Vaccine Undercounter
- Cycle Defrost
- 115v, 60 Hz, 5.4 Amps, 1/3 Hp, Btu/hr 4932
- Ul/c-ul Listed
- Energy Star Certified
- Shipping Weight: 295 Lbs.
- Microprocessor Temperature Controller
- Adjustable Operating Temperature Range: 1°c to 10°c
- Standard Digital Temperature Display and Alarm Module With Battery Back-up to Support Temperature Monitoring and Alarms
- Sample Simulated Digital Temperature Display; .1° Resolution
- Adjustable Temperature Display in Celsius or Fahrenheit
- Audible and Visual High/low Temperature Alarms
- Sensor Error Alarm
- Power Failure Alarm
- Resettable 1-point Min/max History
- Remote Alarm Contacts
- Alarm Validation (Audible, Visual, and Remote Alarm)
- Alarm Mute With Ring Back