View:

BioClave™ Research Autoclaves

  • Designed specifically for the sterilization of laboratory research tools
  • Fully automatic operation – just press start
  • Internal memory for retrieving cycle data 
  • Built in water tank - no external connection needed
  • 8 L, 16 L and 28 L models - all models small enough for benchtop use

 

Despite the space-saving exterior dimensions of the Bioclaves, the stainless steel chambers are surprisingly large, accommodating a variety of liquids, media, instruments, glassware, plasticware, and other common research laboratory items. Available in 8L, 18L, and 28L chamber volumes, these benchtop models include a mechanical and electrical safety interlock that prevents the door from being opened until the pressure has reached zero PSI.

With a large digital LCD display and fully automatic operation, all segments of the cycle (fill, sterilize, exhaust, and dry) commence and run to completion by simply pressing the start button. An optional thermal mini-printer is also available for providing a record of sterilization parameters and a USB drive can be used to record settings and transfer them to a PC.

The 8-liter BioClave Mini is preprogrammed to meet basic sterilization needs. Temperature can be selected at either 121°C or 134°C, and the sterilization time is fully adjustable. The economical BioClave Mini is the ideal choice wherever benchtop sterilization is required and space is limited.

BioClave 18L and 28L Models offer preset cycle options (standard programs), for use with liquids, wrapped research instruments, or unwrapped instruments/plasticware. An additional option “Dry Only” is also available for additional drying time to be added to the end of a cycle. For special applications, requiring variations in cycle parameters, all settings are fully adjustable.

The Bioclave 28L features an increased chamber size with a diameter of 32cm. With this larger internal chamber, the BioClave 28 is our only model capable of sterilizing 1 and 2 liters flasks (and bottles). Despite the increased chamber size, the unit is still relatively compact, suitable for benchtop use.

 

Item#:
ASAUTOC


UV Clave™ Ultraviolet Chamber
  • Reduces potential contaminants
  • Provides true 360° irradiation
  • Proprietary UVC transparent shelf, eliminates potential for missed coverage (blind spots)
  • Inactivates DNA, RNA, viruses and bacteria on PCR tools and vessels
  • Adjustable cycle time

Short-wave UV light is an effective method for the disinfection of surface DNA, RNA and microorganisms. With the UV Clave, highly focused UV irradiation is used to disinfect tools, vessels and other research items, eliminating up to 99% of surface contamination.

Unlike most UV chambers, the UV Clave includes powerful UV-C bulbs on both the top and bottom of the chamber. When combined with our proprietary shelf that allows for >90% of UV-C light transmission and highly reflective chamber walls, the result is true 360° exposure with no blind spots.

Exposure time is adjustable. An optional heating cycle can be added to assist in drying items that may have been washed or chemically disinfected prior to placing them in the chamber.

To protect users, the integrated auto shut-off prevents illumination when the door is open. Optional indicators are available to test for proper exposure levels.


Specifications

UV Bulbs6 x 4 W
UV Wavelength254 nm
UV Intensity+500 uW / cm2 (center of chamber)
UV Dosage600 mJ / cm2 (20 min cycle)
UV Timer5 - 20 min. (5 min. increments)
Heater Timer30 - 120 min. (30 min. increments)
Heater Temp.55°C
Internal Dims.8.5 x 10.1 x 12.2 In. | 21.6 x 25.6 x 31.8 cm
External Dims.11.5 x 10.5 x 15.7 In. | 29.2 x2 6.7 x 40 cm
Max. Height30 cm
Shelf Dims.10 x 8 In. | 25.4 x 20.3 cm
Shelf Material2 Stainless steel wire / 1 UV-C transparent
Weight11 lbs. | 5 kg.
Electrical120 V 60 HZ
Warranty2 Years
Item#:
ASUVCLAVE
View: